Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-12 PerSeptive Biosystems
2
2.3.3 Connecting the
Signatec 500 MHz Digitizer Board
If you have an oscilloscope or an external digitizer on your
system, disregard this section.
CAUTION
Do not use the Signatec 500 MHz digitizer board without
digitizer signal protection circuits (blue boxes on cables).
Operation without digitizer signal protection circuits will
result in damage to the internal digitizer.
To connect the Signatec 500 MHz digitizer board:
Connection on
mass spectrometer
(see Figure 2-2)
Connection on rear panel of
computer (see Figure 2-4)
Cable
TRIG Trig Trig SMA to Trig BNC
CH 1 Channel 1 Ch 1 SMA to Ch 1 BNC
between blue box and
computer.
Ch1 BNC to Ch1 BNC
between blue box and
mass spectrometer.
Ground wire to computer
chassis ground screw.
CH 2
(Voyager-DE PRO
only)
Channel 2 Ch 2 SMA to Ch 2 BNC
between blue box and
computer.
Ch 2 BNC to Ch 2 BNC
between blue box and
mass spectrometer.
Ground wire to computer
chassis ground screw.