Loading Samples on Sample Plates
Voyager
Biospectrometry
Workstation User’s Guide 3-35
3
You can acquire data from a well that does not have an ideal
crystallization pattern. However, when sample contains
unevenly distributed crystalllization, it may be difficult to
analyze. This can cause a problem in Automatic mode.
Other matrices
Typical appearance of other matrices under magnification are
shown in Figure 3-2 and Figure 3-3.
When analyzing 3-HPA crystals, aim the laser at the base of
the fan-like crystals for best response.
Each matrix/sample class may require different laser
intensities for analysis.
Figure 3-2 Microscopic View of Sample Plate with
3-HPA and THAP Matrices
Figure 3-3 Microscopic View of Sample Plate with DHB or
DHBs Matrix
3 HPA
Needles in fan-like arrangement
around outer edge of sample well
THAP
Overlapping shingles
non-uniform shape
PB100508
PB100509
PB100511
PB100510
DHB dried under vacuum
Milky amorphous
DHB or DHBs, air dried
Irregular crystals
in ring