
Chapter 8 Maintenance and Troubleshooting
8-6 PerSeptive Biosystems
8
6. Insert two fuses of the proper rating.
7. Insert the voltage selector/fuse holder into the receptacle.
8. Plug in the mass spectrometer and power up.
8.1.3 Backing Up and Archiving Data
Back up data weekly, or as needed. Archive data as needed.
To operate properly, the Voyager software requires:
• Disk space for data and instrument setting files
• Disk space for printing
If the disk is full, writing data files, printing, or saving
instrument settings can fail and result in loss of data. The
computer also operates more slowly as free disk space
decreases.
To maintain adequate disk space, back up and archive data on
a routine basis:
• Back up—Temporarily store data in case of a hard disk
problem.
• Archive—Permanently store important data for retrieval
at a later time, and remove the original data.
Electrical Rating
Volts/Amps
Fuse
(5 x 20 mm)
100 V~10A T10A 250V
120 V~10A T10A 250V
220 V~6.3A T6.3A 250V
240 V~5A T5A 250V
Comentarios a estos manuales