Voyager DP-263-S Guía de usuario Pagina 267

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 266
Instrument Settings Parameter Descriptions
Voyager
Biospectrometry
Workstation User’s Guide 5-37
5
Accumulation
Conditions
(continued)
Accumulate all passing (continued)—Example: If
Number to Acquire=5, Number of positions in .SP=7,
the number of positions analyzed is determined by
whether acceptance criteria fail:
If all fail, 7 positions analyzed (total number of positions
in .SP), no .DAT created
If the first position fails, but all remaining pass,
2 positions analyzed, one .DAT created
If the first and second positions fail, but all remaining
pass, 3 positions analyzed, one .DAT created, and
so on
If acceptance criteria never fail, 1 search pattern
position analyzed
For more information, see Section 6.6.6, Process that
Occurs when Accumulating Spectra from Multiple Search
Pattern Positions.
Spectrum Acceptance
Criteria
Click to display Spectrum Acceptance Criteria
parameters. See “Setting spectrum acceptance and laser
adjustment criteria” on page 6-40.
* If Use Automated Sample Positioning is disabled, acquisition is performed on the positions you
manually select.
Table 5-10 Automatic Control Parameters—Spectrum Accumulation (Continued)
Parameter Description
Vista de pagina 266
1 2 ... 262 263 264 265 266 267 268 269 270 271 272 ... 611 612

Comentarios a estos manuales

Sin comentarios