Voyager DP-263-S Guía de usuario Pagina 305

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 304
Optimizing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-75
5
5.4.4 Optimizing Signal-to-Noise Ratio
This section includes:
•Overview
Setting Accelerating Voltage
Setting Guide Wire Voltage%
Setting Shots/Spectrum
Setting Low Mass Gate
For more information on measuring Signal-to-Noise, see
Section 6.5.3, Calculating Signal-to-RMS Noise Ratio, and the
Data Explorer Software User’s Guide
, Section 6.5, Using
the Signal-to-Noise Ratio Calculator.
Manually
accumulating
spectra to
improve
signal-to-noise
ratio
The manual accumulation feature of the Voyager system
allows you to acquire a spectrum, examine the spectrum for
signal quality (visually, or using the Resolution or
Signal-to-Noise Ratio calculator), and then accumulate the
spectrum with other previously examined spectra, or discard
the spectrum. This process of accumulating only acceptable
spectra allows you to improve the signal-to noise ratio of the
final spectrum you save. For more information on manually
accumulating spectra, see Section 6.2.2, Manually
Accumulating Spectra from Multiple Acquisitions.
Vista de pagina 304
1 2 ... 300 301 302 303 304 305 306 307 308 309 310 ... 611 612

Comentarios a estos manuales

Sin comentarios