Troubleshooting
Voyager
Biospectrometry
Workstation User’s Guide 8-19
8
8.2.2 Software Troubleshooting
This section includes:
Instrument Control Panel troubleshooting
PSD troubleshooting
Checking the Windows NT Event Log
Table 8-2 Instrument Control Panel Troubleshooting
Symptom Possible Cause Action
Slider control does not
change laser setting
Voyager Instrument
Control Panel is not the
active window
Click on the Instrument
Control Panel to activate
the window before using
slider controls.
Active position number
does not reflect actual
position under laser
Sample plate is not
aligned
Align the sample plate.
See Section 2.7, Aligning
the Sample Plate.
Number of times the laser
fires is greater than
Shots/Spectrum
Delay in transferring data
from the digitizer to the
computer.
No action. Normal
occurrence.
Negative ion selection not
displayed in Instrument
Settings control page (and
you have the optional
hardware)
Instrument not configured
for negative ion hardware
Configure for negative ion
hardware:
1. From the
Mode/Digitizer dialog
box, select the
Instrument Mode tab.
2. Select Negative for
Polarity type and click
OK.