Appendix H Continuous Extraction Mode
H-2 PerSeptive Biosystems
H
H.1 Optimizing a Continuous
Extraction Standard Instrument
(.BIC) Setting
Before you begin
NOTE: Due to the superior results obtained during Delayed
Extraction (DE) mode, use Continuous Extraction mode for
diagnostic purposes only.
Before optimizing a Continuous Extraction .BIC file, be familiar
with the following information:
•
The Data Explorer Software User’s Guide
,
Appendix B, Overview of Isotopes
• “Opening and Viewing Instrument Settings” on page 5-7
• “List of standard instrument settings files” on page H-3
• “Saving and Printing Instrument Settings” on page 5-11
• “Acquiring in Manual Mode from the Instrument Control
Panel” on page 6-11
• Section H.2, Obtaining Good Spectra in Continuous
Extraction Mode
• “Determining Laser Threshold” on page H-12
• Section H.2.3, Checking Resolution
• Section H.2.4, Fine-Tuning the Laser Setting
• Section 5.1, Loading, Modifying, and Saving Instrument
Settings
Instrument
settings
parameters
Most instrument settings parameters are optimized for your
system and should not require adjustment. Change only the
parameters listed in the following procedure.
For more information on remaining instrument settings, see
Section 5.2, Instrument Settings Parameter Descriptions.