Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
Biospectrometry
™
Workstation User’s Guide 6-37
6
Laser
4. Select Use Automated Laser Intensity Adjustment.
NOTE: To manually control the laser intensity when
acquiring in Automatic Control mode, deselect Use
Automated Laser Intensity Adjustment. For more
information, see Section 4.5.2, Adjusting Laser
Intensity and Selecting Sample Position.
5. If you enabled Use Automated Laser Intensity
Adjustment, specify the Minimum and Maximum Laser
Intensity and the Step Size to use. For more information,
see “Automatic Laser Intensity Adjustment” on page 5-32.
6. If you enabled Use Automated Laser Intensity Adjustment
and Use Automated Sample Positioning (described in
step 10 on page 6-39), select PreScan Mode if desired:
• Prescan At First Search Pattern Position Only
• Prescan At Each Search Pattern Position
NOTE: Prescan options are not available unless you
select Use Automated Laser Intensity Adjustment and
Use Automated Sample Positioning.
For more information on the Prescan process, see
“Prescan mode” on page 6-49.
7. If you enabled Use Automated Laser Intensity
Adjustment, set adjustment criteria as described in
“Setting spectrum acceptance and laser adjustment
criteria” on page 6-40.
NOTE: If you do not specify laser adjustment criteria,
the laser is not adjusted. The mid-range laser setting
between the minimum and maximum specified in step 5
is used.