Loading Sample Plates in the Mass Spectrometer
Voyager
™
Biospectrometry
™
Workstation User’s Guide 3-45
3
Figure 3-7 Loading the Sample Plate in a
Voyager-DE STR
WARNING
PHYSICAL INJURY HAZARD. Fingers can get caught in
the sample holder. To avoid injury, do not click Load to
retract the sample holder when your fingers are near the
sample holder.
4. From the Sample Plate menu, select Load to retract the
sample plate and insert it into the main source chamber.
The Load/Eject Sample Plate dialog box (Figure 3-8) is
displayed.
.
Figure 3-8 Load/Eject Sample Plate Dialog Box
PB100277