
Acquiring in Manual Mode from the Instrument Control Panel
Voyager
™
Biospectrometry
™
Workstation User’s Guide 6-13
6
3. Specify calibration information:
• Calibration Matrix—Select the matrix used for
your application. For more information, see “Matrix
influence” on page 5-22.
• Default or External—Select Default, or select an
external .CAL file. For more information, see
“Types of calibration” on page 6-7.
Selecting sample
position and laser
intensity
To select sample position and laser intensity, do the following
in the Manual Laser/Sample Positioning control page (see
Figure 6-2 on page 6-12):
1. If the Manual Laser/Sample Positioning control page is
not displayed, select Manual Laser/Sample Positioning
from the View menu.
2. Select the sample to analyze by doing any of the following
in the Active Position box:
• Type a position number
• Select a position number from the drop-down list
• Single-click on a sample position
Hint: You can also move to a sample position using the
control stick, by selecting Move to Sample Position
from the Sample Plate menu, or by clicking .
To zoom on the Active Sample Position, you can do any
of the following:
• Double-click on the sample position in the Plate
view
• Select Sample View from the Sample Plate menu
• Right-click the mouse and select Toggle
Sample/Plate View.
Comentarios a estos manuales