Voyager DP-263-S Guía de usuario Pagina 395

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 394
Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-83
6
6.7.7.2 Calibration Standard Requirements
Mass calibration
standards
The requirements for mass calibration standards are
determined by your application. The following are general
guidelines:
Mass calibrate on the same sample plate you will use to
analyze samples.
For rapid screening in which high mass accuracy is not
needed, one calibration standard located in the center of
the plate is adequate.
For applications requiring higher mass accuracy, more
calibration standards are needed. You must experiment
to determine the number and location of standards
required to provide the mass accuracy you require.
For optimum mass accuracy, place samples in sample
positions adjacent to standards.
For maximum mass accuracy, spot sample and standard
as close to each other as possible (within 1 mm). This is
referred to as close external calibration. For more
information, see Section 6.7.7.3, Performing Close
External Calibration.
6.7.7.3 Performing Close External Calibration
Overview
You have two options for close external calibration:
Sample and standard in separate sample positions
Sample and standard in the same sample position
NOTE: The closer the proximity of sample and standard,
the better the mass accuracy.
Vista de pagina 394
1 2 ... 390 391 392 393 394 395 396 397 398 399 400 ... 611 612

Comentarios a estos manuales

Sin comentarios