Chapter 6 Acquiring Mass Spectra
6-58 PerSeptive Biosystems
6
6.6.6.2 Process that Occurs when
Accumulating Passing Spectra
When accumulating only the spectra that meet the
Acceptance Criteria (Accumulation mode is determined by the
Spectrum Accumulation conditions described in “Automatic
Control Dialog Box” on page 5-31), the system does the
following during acquisition:
1. Sets the laser intensity as described in Section 6.6.5,
Process that Occurs During Acquisition in Automatic
Mode.
2. Turns on the laser.
3. Acquires the number of Shots/Spectrum at the first search
pattern position.
4. Turns off the laser.
5. Reads the data from the digitizer.
6. Evaluates the data against the Acceptance Criteria in
“Automatic Control Dialog Box” on page 5-31.
7. If the data meets the Acceptance Criteria, holds the data
in memory and averages current data with previous data.
If it fails, the system displays an error message in the
Output window describing which spectrum failed, and
which acceptance criteria specified have failed. The data
is discarded.
8. If the data meets the Acceptance Criteria, stays on the
same position.
If the data fails the Acceptance Criteria, moves to the
next search pattern position.