Appendix H Continuous Extraction Mode
H-18 PerSeptive Biosystems
H
H.2.4 Fine-Tuning the Laser Setting
When you find the laser threshold, whether you need to
fine-tune the setting depends on your needs:
If you are looking for an estimate of molecular
weight, a laser setting slightly higher than the laser
threshold setting may be sufficient.
If you need good peak shapes, you may need to
increase the laser setting to improve signal-to-noise
ratio.
If you need maximum resolution to differentiate
between compounds, you may need to operate
close to the laser threshold setting to improve
resolution.