Appendix A Specifications
A-2 PerSeptive Biosystems
A
A.1 Voyager-DE Specifications
This section includes the following specifications for the
Voyager-DE Workstation:
•Performance
• Mass Spectrometer
• Miscellaneous
Table A-1 Voyager-DE Performance Specifications for Installation
Condition Specification
Resolution (FWHM—Full Width at Half Maximum)
in Delayed Extraction mode
•
≥
50 for IgG (1 pmol/µl)
•
≥
80 for BSA (5 pmol/µl)
•
≥
800 for Insulin, bovine (2 pmol/µl)
•
≥
1,000 for Myoglobin (5 pmol/µl)
•
≥
2,000 for Angiotensin I (1 pmol/µl)
[M+H]
+
ion analyzed.
Mass Accuracy in Delayed Extraction mode
(difference between the mean of six measurements
and the theoretical mass of the sample)
• External Calibration—0.05%
• Internal Calibration—0.02%
Mass accuracy is the difference between the
measured and the theoretical mass of the
sample.
Sensitivity Routine detection of 5 fmol of neurotensin.
Table A-2 Voyager-DE Mass Spectrometer Specifications
Condition Specification
Flight tube 1.2 m linear
Ion source Two-stage