Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-55
6
If Acceptance Criteria are not met, if an accumulation
condition that does not use Acceptance Criteria is
selected, or if the laser can no longer be adjusted, the
system begins a new acquisition from the next search
pattern position, depending on the Spectrum
Accumulation conditions and Spectrum Acceptance
Criteria. Acceptance criteria are defined in “Setting
spectrum acceptance and laser adjustment criteria” on
page 6-40.
If collection mode is set to Save First Passing and
an acceptable spectrum is collected, the system
saves the data file, and the run is complete.
If acceptance criteria are not met, or collection
mode is set to Save all, Save all passing, or Save
best, the acquisition sequence is repeated at the
next position specified in the search pattern file.
7. If the system moves across an entire sample position and
does not find an acceptable signal, it:
Logs the error in the Automatic Control tab in the
Output window of the Instrument Control Panel.
Does not create a data file.
Vista de pagina 366