Impact of Changing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-51
5
Figure 5-12 Effect of Vertical Scale in Digitizer Settings
Suggested settings
Use the settings listed below as a starting point for Vertical
Scale. Adjust as needed to bring the signal into the proper
range.
Vertical
Scale
Decrease if
signal is weak
Increase if
signal exceeds scale
Mass Range (Da) Vertical Scale
0–10,000 1,000 mV
10,000–50,000 200 mV
>50,000 50 mV*
* Not available on the Signatec digitizer.