Voyager DP-263-S Guía de usuario Pagina 168

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 167
Chapter 3 Preparing Samples
3-38 PerSeptive Biosystems
3
CAUTION
Do not leave the sample plate in detergent for longer
than 10 minutes. Longer exposure can cause the
bottom holders on the sample plate to corrode. Do not
sonicate sample plates or use acid to clean sample
plates. Both can alter the surface of the sample plate,
and reduce the quality of the data obtained.
5. If residue remains, wipe the plate with a lint-free tissue or
cotton swab. A soft toothbrush works well.
6. Rinse the plate thoroughly in deionized water.
7. To speed drying, rinse the plate in acetone.
8. Allow the plate to dry in an area where it will not be
exposed to contaminants.
Vista de pagina 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 611 612

Comentarios a estos manuales

Sin comentarios