Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-44 PerSeptive Biosystems
2
Figure 2-26 Sample Plate Alignment Wizard
3. Click Next.
The sample plate moves to the first alignment position
on the sample plate. See Table 2-1, “Four-Corner
Positions on Sample Plates,” on page 2-42, to
determine your plate positions.
4. Start the laser using the control stick.
5. Mark the laser position on the transparency. Do not mark
the sample position. Be sure the laser is centered on the
video monitor and does not appear as shown in
Figure 2-25.
6. Start the laser using the control stick.
7. Center the sample position under the laser using the
control stick.
8. When the position is centered, click Next. The system
advances to the next corner position for the .PLT type.
9. Repeat step 3 through step 8 for each of the four corners
of the sample plate.