Aligning the Sample Plate
Voyager
Biospectrometry
Workstation User’s Guide 2-39
2
2.7 Aligning the Sample Plate
In this section
This section describes:
•Overview
When to align
.PLT files and multiple alignments
How the system aligns a plate
Overview of video monitor display
What you need
Corner positions in .PLT files
Using the control stick
Before aligning
•Aligning
Alignment information saved for future use
Overview
The Sample Plate Alignment function allows you to align the
center of sample positions under the laser beam.
When to align
For most standard welled plates and .PLT files provided,
alignment is not frequently required.
Alignment is more typically required:
When you are using a custom .PLT file and are spotting
sample plates using an automated device.
When you are using Teflon
plates, because there can be
some variation in the laser-etched indentations on the
plate.
Align the Sample Plate if the laser beam is not centered on a
sample position:
When you advance to a sample position using the
Manual Sample Position control page
Before you acquire data from the Instrument
Control Panel and use automated sample
positioning
Before you acquire data from the Sequence
Control Panel