Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
Biospectrometry
™
Workstation User’s Guide 6-57
6
6.6.6.1 Process that Occurs when
Accumulating All Spectra
When accumulating all spectra (Accumulation mode is
determined by the Spectrum Accumulation conditions
described in “Automatic Control Dialog Box” on page 5-31),
the system does the following during acquisition:
1. Sets the laser intensity as described in Section 6.6.5,
Process that Occurs During Acquisition in Automatic
Mode.
2. Turns on the laser.
3. Acquires the number of Shots/Spectrum at the first search
pattern position.
4. Turns off the laser.
5. Holds the data in memory, and averages current data with
previous data.
6. Moves to the next search pattern position.
7. Repeats step 2 through step 6 in subsequent search
pattern positions until any of the following is true:
• Number of spectra to acquire that you select in the
Automatic Control settings is reached
• All search pattern positions have been scanned
• Stop conditions are met
8. Saves the averaged spectrum to disk in the directory
designated in the Data Storage control page, described on
page 6-14.