Aligning the Sample Plate
Voyager
Biospectrometry
Workstation User’s Guide 2-45
2
The software calculates the alignment and uses the
settings to ensure all sample positions are centered
under the laser. If the alignment is successful, a
message is displayed.
10. Do one of the following:
NOTE: A message is displayed if the alignment is outside
the preferred tolerance. Repeat the alignment procedure by
clicking Back. If the alignment is still outside of tolerance,
call PerSeptive Biosystems Technical Support.
Alignment
information saved
for future use
The alignment information is saved with the Plate ID, and
applied each time you load this Plate ID into the system. For
more information, see “How the system aligns a plate” on
page 2-40.
Click To
Finish Save the alignment for the specified
Plate ID
Cancel End the Sample Plate procedure
without saving the alignment for the
plate
Back Repeat the alignment procedure