Voyager DP-263-S Guía de usuario Pagina 190

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 189
Chapter 3 Preparing Samples
3-60 PerSeptive Biosystems
3
Figure 3-12 Area Available for Analysis
How to create a
.PLT file
NOTE: Display the Sample View while creating a .PLT file.
Move the sample plate to the positions you want analyzed
and note the Absolute X, Absolute Y coordinates to enter in
the .PLT file.
To create a .PLT file:
1. Spot a sample plate with
α
-cyano-4-hydroxycinnamic
acid. For information see, “Loading samples” on
page 3-31.
2. Load the sample plate into the system and select
BLANK.PLT as the Plate Type. BLANK.PLT contains one
sample position that occupies the entire plate. For
information, see Section 3.4, Loading Sample Plates
in the Mass Spectrometer.
3. From the View menu in the Instrument Control Panel,
select Manual Laser/Sample Positioning.
Center-to-center distance
X,Y
coordinate
Total
allowable
tolerance
Search pattern
radius
Radius available for analysis
(half the center-to-center distance)
Vista de pagina 189
1 2 ... 185 186 187 188 189 190 191 192 193 194 195 ... 611 612

Comentarios a estos manuales

Sin comentarios