Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-4 PerSeptive Biosystems
2
To select input voltage:
1. Remove the power cord from the mass spectrometer.
2. Carefully remove the voltage selector/fuse holder from the
system (Figure 2-1) using a small flat-blade screw driver.
3. Carefully remove the voltage selector from the holder and
insert the selector with the proper voltage displayed in the
window of the holder.
CAUTION
The plastic tabs that hold the voltage selector in place
are fragile. Do not exert force when removing this
piece.
Figure 2-1 Selecting Input Voltage and Installing
Fuses
PB100507
I
O
240
220
100
120
Fuses
Voltage
selector
Holder
Window