Voyager DP-263-S Guía de usuario Pagina 289

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 288
Optimizing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-59
5
If the spectrum you obtain is not acceptable,
increase or decrease the laser again in 50 to
100-step increments or decrements and reacquire.
If you pass the optimum laser setting, increase or
decrease using a setting that is midway between
the previous two step increments or decrements.
Signal saturation
If the laser intensity is too high, the signal may be saturated
(Figure 5-16). A saturated peak can be determined by:
Figure 5-16 Signal Saturated at High Laser Intensity
Display
Signal is saturated if it
exceeds
Spectrum window 64,000 counts
Oscilloscope (in 1 V full
scale range )
5 screen-grid-line
divisions in the 200 mV
mode
10 screen-grid-line
divisions in the 100 mV
mode
Vista de pagina 288
1 2 ... 284 285 286 287 288 289 290 291 292 293 294 ... 611 612

Comentarios a estos manuales

Sin comentarios