Chapter 7 PSD Analysis
7-28 PerSeptive Biosystems
7
7.3.4 Filling in the Segment List
This section describes:
• Filling in the list
• Typing or selecting new values
• Using the Fill Down command
• Adding or deleting rows
Filling in the list
To fill in the segment list:
If you are using
Angiotensin_PSD.
BIC
1. If you are using the standard instrument settings (.BIC)
file provided, Angiotensin_PSD.BIC, the segment list
contains 10 segments with the Mirror Ratio settings
listed in “Default Mirror Ratio settings” on page 7-22.
If the precursor ion mass you are analyzing differs by
more than 300 Da from the mass in the
Angiotensin_PSD.BIC (1296 Da), you may need to
acquire more or fewer than 10 segments to observe the
fragment ions of interest. You can add or delete
segments, or change the Mirror Ratio for a segment as
needed.
If you are creating
a new .BIC
If you are creating a new .BIC file, click to add the
required number of rows for the number or segments to
acquire. For more information, see Section 7.3.2,
Determining the Number of Segments to Acquire
for a Complete Composite Spectrum.