Acquiring PSD Data with Standard Instrument Settings (.BIC) Files
Voyager
Biospectrometry
Workstation User’s Guide 7-31
7
NOTE: Do not select Guide Wire% Tracks Mirror
Ratio option on Voyager-DE PRO or Voyager-DE
STR systems. This parameter is for use with older
systems only.
2. Type in values in any row that you want to fill down into
selected rows.
3. Click on the Segment number to select the row
containing the values to fill down.
4. Click . The following occurs:
The Mirror Ratio is calculated and entered in all
rows below. If no Decrement Ratio is specified, the
value from the first cell is copied to selected cells
below.
NOTE: If you selected Guide Wire% Tracks Mirror
Ratio, Guide Wire% is also calculated and filled in.
The information in the first row of remaining
columns is copied to all rows below.
The Max Stitch Mass is calculated and displayed
for each row.
Any rows for segments that have already been
saved are skipped.
Adding or
deleting rows
To add rows, select any existing row then click . A row is
added to the end of the list and the settings from the selected
row (or the first row if no row is selected) are copied to the
added row. Change settings as needed.
To delete rows, select the rows to delete and click . You
cannot delete a row if the segment has been saved.