Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-40 PerSeptive Biosystems
2
.PLT files and
multiple
alignments
A .PLT file defines the sample positions on a sample plate. For
example, 100.PLT may define a 100-well plate. If you have
more than one 100-well plate, you may need to align each
plate. The Voyager software allows you to assign a unique
Plate ID to each plate that allows customized alignment of
more than one plate that uses the sample .PLT file
(Figure 2-22).
Figure 2-22 Multiple Alignments Using Sample .PLT
File
How the system
aligns a plate
When you load a sample plate, the system aligns the plate as
described below:
If the loaded plate has a Plate ID, the software looks for
alignment information associated with the Plate ID and
adjusts accordingly.
If there is no alignment information available for the Plate
ID (sample plate has not been aligned), the software uses
the default alignment.
Plate A
100well.PLT
Aligned
6/1/99
Plate ID 1A
Plate B
100well.PLT
Aligned
6/15/99
Plate ID 1B
Plate C
100well.PLT
Aligned
6/30/99
Plate ID 1C