Voyager DP-263-S Guía de usuario Pagina 320

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 319
Chapter 6 Acquiring Mass Spectra
6-8 PerSeptive Biosystems
6
Sequence Control calibration–Provides external,
internal, and internal with automatic updating calibration
options during acquisition from the Sequence Control
Panel. For more information, see Section 6.7.7.1,
Calibration Options in a Sequence.
For the mass accuracy specifications for your instrument, refer
to Appendix A, Specifications.
NOTE: Mass accuracy for sample positions in the outer
rows of a sample plate may be poorer than the mass
accuracy of the sample positions in the inner rows.
When to calibrate
Calibrate:
At least once a day. You may want to include calibration
standards on each sample plate.
If the ambient temperature in your lab is fluctuating
more than 5°C, check calibration, and recalibrate if
necessary.
The calibration software compensates for changes in
instrument settings. However, if you require optimum mass
accuracy, generate a calibration file with identical instrument
settings or use internal calibration if possible.
Vista de pagina 319
1 2 ... 315 316 317 318 319 320 321 322 323 324 325 ... 611 612

Comentarios a estos manuales

Sin comentarios