Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-39
6
Sample positioning
10. Select Use Automated Sample Positioning.
NOTE: To manually adjust sample positioning when
acquiring in Automatic Control mode, deselect Use
Automated Sample Positioning. For more information,
see Section 4.5.2, Adjusting Laser Intensity and
Selecting Sample Position.
11. If you enabled Use Automated Sample Positioning, select
a search pattern (.SP) file. The number of positions in the
selected .SP file is displayed.
For information on search pattern files, see Section
6.6.4, Search Pattern Files.
Stop conditions
12. Enable Stop Conditions if you want to stop the
acquisition when:
Specified number of consecutive acquisitions fail
acceptance criteria
Specified number of laser adjustments within a
search pattern position do not yield acceptable
data when using manual sample positioning
For more information, see “Stop Conditions” on
page 5-38.
13. Click OK.
For more information on Automatic Control parameters,
see Section 5.2.3, Automatic Control Dialog Box.