Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
Biospectrometry
™
Workstation User’s Guide 6-35
6
See Section 2.7, Aligning the Sample Plate, to determine if
sample plate alignment is necessary. If you do need to align
the sample plate, you must do so for each sample plate you
use.
6.6.2 Setting Instrument Settings for
Automatic Control Mode
This section includes:
• Setting Automatic Control settings
• Setting spectrum acceptance and laser adjustment
criteria
• Saving the instrument settings (.BIC) file
Setting Automatic
Control settings
After you optimize an instrument settings file for Manual
Control mode and ensure that parameters are adjusted to
yield optimum performance, you can adjust the instrument
settings for Automatic Control mode, if desired. For
information on optimizing for Manual Control mode, see
Chapter 5, Optimizing Instrument Settings.
You can set parameters to control the following in Automatic
Control mode:
• Laser intensity adjustment
• Spectrum accumulation and saving
• Sample positioning
• Stop conditions
To set instrument settings for Automatic Control mode:
1. In the Instrument Control Panel, open or create an
instrument settings (.BIC) file with the appropriate
parameters. For more information, see Section 5.4,
Optimizing Instrument Settings Parameters.