Aligning the Sample Plate
Voyager
Biospectrometry
Workstation User’s Guide 2-43
2
Using the control
stick
To align a sample plate, you must use the control stick. For
details on using the control stick, see Appendix I, Using the
Oscilloscope and Control Stick.
Before aligning
Before aligning the sample plate:
1. Spot the sample plate with matrix as described in “What
you need” on page 2-42.
2. Load the sample plate into the system and select the
Plate ID that corresponds to the plate:
From the Sample Plate menu, select Select
Sample Plate to display the Select Sample Plate
Type dialog box.
Select a Plate ID. The .PLT file and alignment
information associated with the Plate ID are
automatically loaded.
Alternatively you can specify a new Plate ID and
select a .PLT file. For more information, see
“Assigning Plate IDs” on page 3-39.
Click OK.
Aligning
To align the loaded plate:
1. Tape the transparency to the video monitor.
2. Select Align Sample Plate from the Sample Plate menu.
The Sample Plate Alignment Wizard dialog box
(Figure 2-26) is displayed.