Selecting the Site
Voyager
Biospectrometry
Workstation User’s Guide 2-5
2
WARNING
FIRE HAZARD. Using a fuse of the wrong type or rating
can cause a fire. Replace fuses with those of the same
type and rating.
4. Insert two fuses of the proper rating for the selected
voltage.
5. Insert the voltage selector/fuse holder into the receptacle.
6. Plug in the mass spectrometer. Connect devices to the
mass spectrometer as described in Section 2.3,
Connecting Voyager-DE and Voyager-DE PRO
Workstations, then power up.
Electrical Rating
Volts/Amps
Fuse
(5 x 20 mm)
100 V~10A T10A 250V
120 V~10A T10A 250V
220 V~6.3A T6.3A 250V
240 V~5A T5A 250V