Troubleshooting in Continuous Extraction Mode
Voyager
Biospectrometry
Workstation User’s Guide H-21
H
Poor resolution in
Continuous Extraction
mode
(continued on next page)
Laser intensity too high Adjust laser by using the
slider controls on the
Manual Laser Control
page.
Accelerating Voltage
incorrect
Adjust.
Guide Wire Voltage% too
high
Adjust. See Section 5.3.4,
Understanding Guide Wire
Voltage%.
Current calibration not
correct for sample (off by
a factor of 2 or 3)
Use correct calibration for
sample.
Sample concentration too
high
Prepare sample/matrix
with a final sample
concentration appropriate
for sample and matrix.
See “Matrix Information”
on page 3-6.
Excess matrix in sample
preparation
Increase sample:matrix
ratio.
Too much salt or buffer in
sample
Clean up sample. See
Section 3.1.5, Sample
Cleanup.
On Voyager-DE PRO and
Voyager-DE STR
systems, not separating
isotopic peaks in reflector
mode up to mass 1,000
Adjust Accelerating
Voltage to 10,000 V and
reacquire spectrum. See
Section 5.4.4.2, Setting
Accelerating Voltage.
Table H-8 Spectrum Troubleshooting (Continued)
Symptom Possible Cause Action