Chapter 8 Maintenance and Troubleshooting
8-4 PerSeptive Biosystems
8
Inspect flap valve 1, flap valve 2, linear actuator o-rings
Clean optics and air lines
Inspect grids, compressor, air pressure, turbo pump, and
laser power
Adjust laser flash rate, load offsets, sample offsets,
detector gain and voltage, and instrument covers
Inspect load and eject cycles, high voltage power
supplies, and computer
Calibrate thermocouple gauges
Check that the instrument meets specifications
Changing fuses
This procedure is required for Voyager-DE and
Voyager-DE PRO systems only (the Voyager-DE STR does
not require power fuses). Extra fuses for different voltage
settings are supplied with the system.
To change the main power fuse:
1. Exit the Voyager software.
2. Power down the mass spectrometer.
3. Remove the power cord from the mass spectrometer.
4. Carefully remove the voltage selector/fuse holder from the
system (Figure 8-1) using a small flat-blade screw driver.
5. Carefully remove the voltage selector from the holder and
insert the selector with the proper voltage displayed in the
window of the holder.
CAUTION
The plastic tabs that hold the voltage selector in place are
fragile. Do not exert force when removing this piece.