
Chapter 4 Voyager Instrument Control Panel Basics
4-12 PerSeptive Biosystems
4
7. Set as needed:
8. Click OK.
Parameter Description
Scaling Mode
Display Relative Autoscales the trace to the largest peak in the selected range.
Base Peak
Relative
Autoscales the trace to the base peak in the entire range, not the
selected range. Displays a right axis label with the base peak
intensity.
NOTE: To turn off the right axis, select Graphic Options from
the Display menu, click the Graph #1 Setup tab, and deselect
the Show Right Axis check box.
Absolute Value Autoscales the trace to the Y Display Range entered.
Minimum Absolute Max Y
Use Limit Sets the minimum value for Y scaling. Useful to maintain relative
scaling of peaks and to prevent autoscaling noise to full scale.
Y Display Range
From/To Sets the range for scaling. When Scale To Display Window or
Scale to Base Peak is selected, range is in percent.
When Absolute Value is selected, range is in counts.
Current Trace
Min/Max
Resets range to the default range for the current trace.
Comentarios a estos manuales