Voyager DP-263-S Guía de usuario Pagina 240

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 239
Chapter 5 Optimizing Instrument Settings
5-10 PerSeptive Biosystems
5
3. Select Manual Control mode. For parameter descriptions,
see Section 5.2.1, Instrument Settings Page.
4. Adjust the mass range if needed.
5. To include matrix peaks in the spectrum for calibration,
deselect Low Mass Gate and set the mass to a mass
below the matrix peak mass. For matrix masses, see,
Appendix C, Matrices.
6. Select a calibration file (.CAL), or if you are screening
samples, use the default calibration. For better mass
accuracy, select a .CAL file you recently (same day)
generated in the Data Explorer software. For more
information, see the
Data Explorer Software User’s
Guide
, Section 5.3.2, Manually Calibrating.
7. Save the .BIC file. For more information, see
Section 5.1.5, Saving and Printing Instrument Settings.
Optimizing
For suggested strategies for optimizing parameters, see:
Section 5.3, Impact of Changing Instrument Settings
Parameters
Section 5.4, Optimizing Instrument Settings Parameters
Modifying for
Automatic
Control mode
After you optimize the instrument settings file for Manual
Control mode and ensure that parameters are adjusted to
yield optimum performance, you can adjust the instrument
settings for Automatic Control mode, if desired. For more
information, see Section 6.6.2, Setting Instrument Settings for
Automatic Control Mode.
Vista de pagina 239
1 2 ... 235 236 237 238 239 240 241 242 243 244 245 ... 611 612

Comentarios a estos manuales

Sin comentarios