Appendix H Continuous Extraction Mode
H-20 PerSeptive Biosystems
H
Refer to the table below if you are having spectrum trouble:
Signal is flat Laser setting may be too low, increase.
Sample may be consumed, move to a new position
in sample well.
Sample may not be present, try new position.
Sample may not ionize well, use different matrix.
Cannot see ions in Reflector
mode
Check that you can see ions in Linear mode:
If you can see ions in Linear mode, it indicates
that voltages or laser power need adjusting.
If you cannot see ions in Linear mode, refer to
Section 8.2, Troubleshooting.
Table H-7 Laser Threshold Troubleshooting (Continued)
Symptom Action
Table H-8 Spectrum Troubleshooting
Symptom Possible Cause Action
Very wide peaks Refer to “Poor resolution
in Continuous Extraction
mode” symptom on
page H-21
Refer to “Poor resolution
in Continuous Extraction
mode” action on
page H-21.