Voyager DP-263-S Guía de usuario Pagina 172

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 171
Chapter 3 Preparing Samples
3-42 PerSeptive Biosystems
3
3.4.2 Loading Sample Plates
This section describes loading sample plates in:
Voyager-DE and Voyager-DE PRO
Voyager-DE STR
Voyager-DE and
Voyager-DE PRO
To load sample plates:
1. Eject the sample holder as described in Section 3.4.1,
Ejecting the Sample Holder.
2. Hold the sample plate with the bottom of the numbers
facing toward the analyzer (for standard 100-well plate)
and with the slanted underside of the plate facing to the
left.
3. Slide the sample plate into the holder from the right side
until it snaps into place (Figure 3-5). The ball bearings on
the holder snap into the plate.
CAUTION
If the sample plate does not snap into place, it may be
inserted into the holder the wrong way, and it may jam
inside the instrument. Remove the plate, slide it into the
holder with the slanted underside of the plate facing to
the left and toward the back of the instrument, and snap
it into place.
Vista de pagina 171
1 2 ... 167 168 169 170 171 172 173 174 175 176 177 ... 611 612

Comentarios a estos manuales

Sin comentarios