Chapter 6 Acquiring Mass Spectra
6-16 PerSeptive Biosystems
6
During
acquisition
During acquisition:
The Live/Current Spectrum trace in the Spectrum window
updates to display the spectrum that results from each
laser shot.
NOTE: If your system includes an oscilloscope, the
Current Spectrum does not display a trace until
acquisition is complete. The spectra that result from
each laser shot are displayed on the oscilloscope
screen.
The system averages all spectra acquired since you
started acquisition.
When acquisition is complete, the software displays the
averaged spectrum in a Current trace in the Spectrum window.
Evaluating data
As data is acquired, observe the trace in the Spectrum
window:
1. Examine the number at the top of the right axis in the
Current Spectrum trace to determine if signal is
saturated (Figure 6-4). Signal with counts greater than
64,000 may be saturated. Adjust the laser intensity as
needed.
For information, see “Signal saturation” on page 5-59.
2. Use the Resolution calculator or the Signal-To-Noise
calculator on the Current trace to determine if data is
acceptable. For more information see Section 6.5,
Evaluating Data in the Instrument Control Panel.
Hint: You can accumulate traces to improve resolution
and signal-to-noise ratio. For more information, see
Section 6.2.2, Manually Accumulating Spectra
from Multiple Acquisitions.