Chapter 2 Installing the Voyager™ Biospectrometry™ Workstations
2-32 PerSeptive Biosystems
2
• Source Chamber (BA1) Wait Time—Time
(seconds) that the software waits for the
instrument to reach the Source Chamber
Maximum Operating Pressure, after the sample
plate is loaded. If the wait time is exceeded, an
error message is displayed which gives you the
option of an additional wait time or ejecting the
plate. Valid range is 0 to 300 seconds. Default is
120 seconds.
• Sample Loading Chamber (TC2) Max Load
Pressure—Pressure (Torr) which must be
reached before the system moves the sample
plate from the load chamber to the source
chamber during the load cycle. Valid range is
8x10
-2
to 10
-3
. Default is 8x10
-2
. If Max Load
Pressure is not reached, sample plate will not
load.
• Sample Loading Chamber (TC2) Wait Time—
Time (seconds) that the software waits for the
instrument to reach the Sample Loading Chamber
Maximum Load Pressure after the sample plate is
loaded. If the wait time is exceeded, an error
message is displayed which gives you the option
of an additional wait time or ejecting the plate.
Valid range is 0 to 300 seconds. Default is
120 seconds.
4. Click OK to exit vacuum configuration.