Acquiring PSD Data with Standard Instrument Settings (.BIC) Files
Voyager
Biospectrometry
Workstation User’s Guide 7-17
7
7.3 Acquiring PSD Data
with Standard Instrument
Settings (.BIC) Files
In this section
This section includes:
Steps in acquiring PSD data
Practicing in PSD mode
Differences from regular analysis
Steps in acquiring
PSD data
The steps in acquiring PSD data include:
Practicing in PSD
mode
This section tells you how to use the software to collect
spectra in PSD mode. However, to be successful in PSD
analysis, you need to understand how spectra behave in PSD
mode, and how to optimize acquisition conditions.
Follow the steps in this section to understand how to use the
software. Then follow the steps in Section 7.4, Exploring PSD
Mode, to learn how to optimize the parameters that impact
PSD analysis.
NOTE: Before running unknowns, we recommend
performing the steps above using a standard.
Step Described on
Determining the precursor ion mass page 7-19
Determining the number of segments to acquire for a complete
composite spectrum
page 7-21
Setting the PSD Acquisition parameters page 7-24
Filling in the segment list page 7-28
Acquiring and saving PSD segments page 7-32