Instrument Settings Parameter Descriptions
Voyager
Biospectrometry
Workstation User’s Guide 5-19
5
Shots/Spectrum Determines the number of laser shots that each spectrum
will contain. For more information, see Section 5.4.4.4,
Setting Shots/Spectrum.
The maximum number of Shots per Spectrum for each
digitizer are as follows:
Signatec—256
LeCroy—1,000
Tektronix—10,000
NOTE: The actual number of times the laser fires may be
greater than the number of Shots/Spectrum specified, due
to the limited averaging speed of the digitizer. For
example, if 250,000 data points are recorded, every other
laser shot is skipped by the digitizer. In any case, the
specified number of Shots/Spectrum will be stored in the
data file.
Spectrum Acquisition
Mass Range (Da) Determines the mass range collected.
Valid ranges are:
Start Mass—0 to End Mass.
End Mass—Start Mass to End Mass.
For information on the default digitizer settings used with
the mass range specified, see “Default Number of Data
Points/Bin Size settings” on page 5-48.
Table 5-4 Instrument Settings Parameters (Continued)
Parameter Description