Voyager DP-263-S Guía de usuario Pagina 308

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 612
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 307
Chapter 5 Optimizing Instrument Settings
5-78 PerSeptive Biosystems
5
A lower Accelerating Voltage setting does the following:
Provides more data points across a peak for better peak
definition when analyzing low mass ions (Linear mode
only).
Improves resolution when analyzing compounds below
2,000 Da, and the resolution is limited by the digitization
rate of the system (Reflector mode).
5.4.4.3 Setting Guide Wire Voltage%
To optimize sensitivity, you can adjust the Guide Wire
Voltage%:
Linear mode—Increase in increments of 0.05 percent
until you see signal-to-noise ratio start to decrease.
Reflector mode—Increase in increments of
0.01 percent. Settings above 0.05 percent do not typically
improve signal-to-noise ratio.
NOTE: In Reflector mode, Guide Wire Voltage% also affects
mass resolution. To fine tune signal-to-noise ratio and
resolution at the same time, adjust the Guide Wire Voltage%
in increments smaller than 0.01 percent.
For more
information
For more information, see:
Section 5.4.3.3, Optimizing Guide Wire Voltage%
Section 5.3.4, Understanding Guide Wire Voltage%
Vista de pagina 307
1 2 ... 303 304 305 306 307 308 309 310 311 312 313 ... 611 612

Comentarios a estos manuales

Sin comentarios