Chapter 6 Acquiring Mass Spectra
6-34 PerSeptive Biosystems
6
6.6 Acquiring in
Automatic Mode from the
Instrument Control Panel
This section includes:
• Before acquiring in Automatic Control mode
• Setting Instrument Settings for Automatic Control mode
• Automatically acquiring, evaluating, and saving spectra
• Search Pattern files
• Process that occurs during acquisition in Automatic
Control mode
• Process that occurs when accumulating spectra from
multiple search pattern positions
6.6.1 Before Acquiring in
Automatic Control Mode
Background
information
Before acquiring spectra in Automatic Control mode, become
familiar with the information in:
• Section 6.1, Before You Begin
• Section 6.2, Acquiring in Manual Mode from the
Instrument Control Panel
Aligning the
sample plate
Sample plate alignment is necessary for automated sample
positioning in Automatic Control mode if the laser is not
striking the center position of the sample position.
Sample plate alignment may not be necessary on your
system, particularly if you use the SPIRAL.SP search pattern
file provided with the system.