Chapter 6 Acquiring Mass Spectra
6-84 PerSeptive Biosystems
6
Sample and
standard in
separate sample
positions
If you are acquiring samples and standards from different
sample positions, enter the standards in the run list preceding
the unknowns that use the calibration.
For optimum mass accuracy, place samples in sample
positions adjacent to standards. See the example in
Figure 6-24.
Figure 6-24 Sample and Standard Location for
Optimum Mass Accuracy
To run samples spotted as illustrated in Figure 6-24, set up the
run list as shown in the following table.
= Empty
44 45 46
54 55 56
= Standard
34 35 36
43
53
33
47
57
37
= Sample
Position Base File
Name
Calibration Type Calibration
File
45 STD1 INTERNAL-UPDATE STD.CAL
34 SAMP1 EXTERNAL STD.CAL
35 SAMP2 EXTERNAL STD.CAL
36 SAMP3 EXTERNAL STD.CAL
44 SAMP4 EXTERNAL STD.CAL
46 SAMP5 EXTERNAL STD.CAL
54...56 SAMP6...8 EXTERNAL STD.CAL