Loading Samples on Sample Plates
Voyager
Biospectrometry
Workstation User’s Guide 3-33
3
3.2.3 Loading Samples
(Thin Layer Application)
When to use
Use the thin layer application technique
1
for increased
sensitivity when analyzing peptides with a concentration
<0.1 pmol/ µl.
Loading matrix
and samples
To load matrix and samples:
1. Dispense 0.5 µl of matrix prepared for thin layer
application on the sample plate to form a thin layer. It
will dry in seconds.
For information, see “CHCA for thin layer application”
on page 3-9.
2. Load 0.5 to 1 µl of the sample in each sample position
onto the thin matrix layer applied in step 1.
3. Dry the plate with a heat gun.
4. After drying, load 2 to 3 µl of double-distilled water on
top of each dried sample to wash the sample.
5. Wait 5 seconds, and blow off the water droplet with
compressed air.
6. Repeat step 4 and step 5 two more times.
NOTE: During acquisition, use the “surfing method”
(moving the sample plate after every 5 to 10 shots, to a new
position on the sample).
1. Shevchenko, A., M. Wilm, O. Vorm, M. Mann,
Anal. Chem.,
1996, 68,
850–858.