Chapter 7 PSD Analysis
7-2 PerSeptive Biosystems
7
7.1 Overview of PSD Analysis
This section includes:
Post-source decay analysis
Segments and composite spectra
PSD data files
Mass calculation for fragment ions
Optimizing the Precursor Ion Selector
NOTE: Analysis of post-source decay is available on
Voyager-DE PRO and Voyager-DE STR workstations only.
Before you begin
For successful analysis of post-source decay, you must be
familiar with the information in:
Chapter 5, Optimizing Instrument Settings
Chapter 6, Acquiring Mass Spectra
7.1.1 Post-Source Decay Analysis
What is
post-source
decay analysis?
You can obtain valuable structural information about a sample
by analyzing the fragment ions generated from the original
ions in the flight tube. On a MALDI-TOF system, this is known
as post-source decay (PSD) analysis.
PSD analysis on the Voyager system is an acquisition method
tailored to the analysis of fragment ions. PSD mode allows you
to better analyze ions that fragment in the flight tube by
optimizing the following settings:
Mirror Ratio setting—Brings different mass ranges of
fragment ions into focus.
Precursor Ion Selector—Selectively analyzes the ion of
interest, without interference from other compounds.
Specificity of ion selection is determined by the resolution
of the Precursor Ion Selector on your system. See
Appendix A, Specifications, for more information.