Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-77
6
6.7.4 Preparing to Run a Sequence
This section describes:
Aligning the sample plate
High voltage warmup for improved mass accuracy
Before acquiring
Aligning the
sample plate
Sample plate alignment is necessary for a sequence run if the
laser is not striking the center of the sample position.
Sample plate alignment may not be necessary on your
system, particularly if you use 100-well plates and the
SPIRAL.SP search pattern file provided with the system.
See Section 2.7, Aligning the Sample Plate, to determine if
sample plate alignment is necessary. If you do need to align
the sample plate, you must do so for each sample plate you
use.
High voltage
warmup for
improved mass
accuracy
For maximum mass accuracy, allow the high voltage power
supplies to warm up for a short period of time before
acquisition. This reduces variability in accelerating voltages,
and yields more reproducible ion flight times.
To turn on the high voltage power supplies, click in the
toolbar.
NOTE: The high voltage power supplies are automatically
turned off after 60 minutes if the mass spectrometer is not
used. To change the Idle Time see “High voltage
configuration” on page 2-33.