Instrument Settings Parameter Descriptions
Voyager
Biospectrometry
Workstation User’s Guide 5-27
5
Table 5-7 Linear and Reflector Digitizer Parameters
Parameter Description
Horizontal Settings Optimization
Bin Size (nsec) Determines the time (nanoseconds) interval between
subsequent data points (see Figure 5-11 on page 5-49). Use
this setting to optimize resolution. Bin size and Number of
Data Points are dependent values. The Bin sizes available on
your system depend on the frequency (500 MHz to 4 GHz) of
your digitizer.
By default, the system calculates a Bin size that is compatible
with the mass range you specify and that yields 100,000 data
points. For information on overriding this value, see
“Overriding horizontal digitizer settings” on page 5-48.
For more information, see Section 5.3.5, Understanding
Digitizer Settings.
Number of Data
Points
Selects number of data points. Use this setting to optimize file
size. Bin size and Number of Data Points are dependent
values. The Number Of Data Points available on your system
depend on the frequency (500 MHz to 4 GHz) of your
digitizer.
By default, Number of Data Points is set to 100,000. For
information on overriding this value, see “Overriding
horizontal digitizer settings” on page 5-48.
For more information, see Section 5.3.5, Understanding
Digitizer Settings.