Chapter 5 Optimizing Instrument Settings
5-4 PerSeptive Biosystems
5
List of .BIC files
Table 5-1 through Table 5-3 list the standard .BIC files
provided on your system for the following modes:
Linear mode
Reflector mode
PSD mode
Table 5-1 Linear Mode .BIC Files
.BIC File Sample Test
Mass Range in
.BIC Optimized
at (Da)*
Angiotensin_Linear.BIC Low mass
peptide mix1
Calibration and
Resolution
(angiotensin I)
500–2,000
ACTH_Linear.BIC Peptide mix2 Calibration and
Resolution
(ACTH 18–39)
2,000–3,000
Insulin_Linear.BIC
Peptide mix2 Resolution
(insulin)
5,000-6,000
Myoglobin_Linear.BIC Low mass
protein mix3
Resolution
(myoglobin)
15,000–20,000
BSA_Linear.BIC BSA Resolution 60,000–70,000
IgG_Linear.BIC IgG Resolution 100,000–
200,000
Peptide_Sensitivity_Linear.BIC Neurotensin
in mix1
Sensitivity 1,000–2,000
Peptide_Linear_Negative.BIC Low mass
peptide mix1
Negative ion
mode
500–2,000
* Mass Range specified for acquisition may be wider.